User's Manual

Table Of Contents
18 - 5
18.2.6.3 Editing the Graphic Trend Group
1. Enter the graphic trends review page.
2. Select
Group Setup.
For more information, see
18.2.5.3Editing the Tabular Trend Group for detail on how to edit the group.
18.2.6.4 Changing the Resolution of Trend Data
To change the length of trend data displayed on the current screen, follow this procedure:
1. Enter the graphic trends review page.
2. Select
Zoom.
8 min: the screen displays eight minutes of trend data. You can view the most recent one hour of
data.
30 min, 1 hr, 2 hrs, 4 hrs: the screen displays 30 minutes, one hour, two hours, or four hours of trend
data. You can view the most recent four hours of data.
8 hrs, 12 hrs, 24 hrs, 48 hrs: the screen displays 8 hours, 12 hours, 24 hours, or 48 hours of trend data.
You can view the most recent 120 hours of data.
18.2.6.5 Changing the Number of Waveforms
To change the number of waveforms displayed on the trend review page, follow this procedure:
1. Enter the graphic trends review page.
2. Select
Trends.
18.2.6.6 Printing a Graphic Trends Report
To print a graphic trends report, follow this procedure:
1. Enter the graphic trends review page.
2. Select
to enter the Print Setup dialog.
3. Set the graphic trends report as described in
22.6.4Configuring Graphic Trends Reports.
4. Select
Print to print the graphic trend data before the above set Time and for the above set Period.
18.2.7 Events Review Page
The monitor stores events in real time, including technical alarm events, physiological alarm events, manual
events, and operational events. When an event occurs, all the measurement numerics and three event-related
NOTE
waveforms 16 seconds before and after the event are stored.
Alarms are saved as events and will be maintained if the equipment is powered down. The time of
equipment power down is not recorded as an event and cannot be reviewed.
Earlier events will be overwritten by later ones if the storage capacity is reached.
A total loss of power does not affect the events already stored.
18.2.7.1 Entering the Events Review Page
Choose one of the following methods to enter the events review page:
Select the Review quick key select the Events tab.
Select the Main Menu quick key from the Review column select Events.
The
Event page displays the event list. Events are displayed in descending chronological order, with the most
recent displayed at the top. The number of asterisk symbols before an event indicate alarm priority as described
in
7.3.3Alarm Indicators.
Different color blocks are displayed on the left of each event to indicate different event types.