User's Manual

Table Of Contents
3 - 1
3 Getting Started
3.1 Equipment Preparation Safety Information
WARNING
Use only installation accessories specified by Mindray.
Connect only approved devices to this equipment. Devices connected to the equipment must meet
the requirements of the applicable IEC standards (e.g. IEC 60950 safety standards for information
technology equipment and IEC 60601-1 safety standards for medical electrical equipment). The
system configuration must meet the requirements of the IEC 60601-1 medical electrical systems
standard. Any personnel who connect devices to the equipments signal input/output port are
responsible for providing evidence that the safety certification of the devices has been performed in
accordance to the IEC 60601-1. If you have any questions, please contact Mindray.
If it is not evident from the equipment specifications whether a particular combination with other
devices is hazardous, for example, due to summation of leakage currents, please consult the
manufacturer or an expert in the field. A determination must be made that the proposed
combination will not negatively affect the devices themselves or the patient's safety.
CAUTION
The equipment should be installed by Mindray authorized personnel.
When disposing of the packaging material, be sure to observe the applicable waste control
regulations and keep it out of childrens reach.
Before use, verify whether the packages are intact, especially the packages of single use accessories.
In case of any damage, do not apply it to the patient.
Make sure that the equipment operating environment meets the specific requirements. Otherwise
unexpected consequences, e.g. damage to the equipment, could result.
Avoid rough handling during transport.
NOTE
Put the equipment in a location where it can be easily viewed and operated.
Keep this manual in the vicinity of the equipment so that it can be referenced when needed.
Save the packing case and packaging material as they can be used if the equipment must be
reshipped.
Many settings in the patient monitor are password protected. It is recommended to change the
default passwords and keep the passwords safe. Passwords should only be changed by authorized
personnel. Contact the department manager or biomedical engineering department for the
passwords used at the facility.
3.2 Monitor Installation
The monitor can be installed in various ways as required.
Wall mount
Placed on desk
Trolley tray
Bedrail clamp
Bedrail hook