If the sidestream CO
2
module is configured with the oxygen sensor, O
2
waveform and parameters can be
displayed in an O
2
numeric and waveform area as follows:
16.6 Measuring CO
2
Using the Sidestream/Microstream CO
2
Module
16.6.1 Preparing to Measure CO
2
Using the Sidestream CO
2
Module
To prepare the CO
2
module for measurement, follow this procedure:
1. Select the appropriate gas sample line and watertrap according to the patient category.
2. Connect the watertrap to the CO
2
module, and connect the gas sample line to the watertrap.
3. Connect the other end of the gas sample line to the patient.
For intubated patients requiring an airway adapter, install the airway adapter between the patient
circuit and the ventilator Y-piece.
(1) O
2
waveform (2) FiO
2
measurement (3) EtO
2
measurement
(3)
(2)
(1)
(1) Watertrap receptacle (2) DRYLINE II watertrap (3) Gas sample line
(1) Sample line (2) Connection to the ventilator
(3) Airway adapter (4) Connection to the patient
(1)
(2)
(3)
(1)
(2)
(3)
(4)