C.O. measurement fails 1. Inject more volume, or inject colder solution. Make sure that the
injectate temperature is at least 10°C colder than the patient blood
temperature.
2. Finish injection within four to five seconds.
3. Check the connection of C.O. cable, thermodilution catheter and TI
sensor.
Problem Solution