14 - 11
WARNING
Refer to the Camino Operator’s Manual to adjust settings and connect the monitor with the patient.
ICP alarm settings on the Mindray patient monitor are independent of the Camino monitor alarm
settings, and thus may be different. Please pay special attention to the alarms on the Camino
monitor.
14.9 IBP Troubleshooting
This section lists the problems that might occur. If you encounter the problems when using the equipment or
accessories, check the table below before requesting service. If the problem persists, contact your service
personnel.
NOTE
For the physiological and technical alarm messages, see Appendix DAlarm Messages.
(1) External IBP module (2) Camino ICP cable
(3) ICP adapter (4) ICP connector
(5) Camino cable (6) Camino monitor
Problem Solution
Cannot see IBP numeric area or waveform
area on the main screen
1. Check that the IBP is set to display in the Screen Setup dialog. For more
information, see.
3.9.2Displaying Parameter Numerics and Waveforms.
2. Check that if the IBP parameter switch is enabled. If not, enable the IBP
measurement. For more information, see
3.9.1Switching On or Off a
Parameter
.
3. Check the connection of IBP cable, IBP transducer and the monitor.
4. Check that the three-way valve is turned to the correct position.
5. Check that the IBP transducer has been zeroed. For more information, see
14.3.3Zeroing the IBP transducer.
Cannot see systolic pressure and diastolic
pressure for P1/P2/P3/P4
Set
Measure to All in the P1/P2/P3/P4 setup page. For more information, see
14.6.3Setting the Pressure Type for Display.
IBP readings seem unstable 1. Make sure there are no air bubbles in the transducer systems.
2. Check that the transducer is properly fixed.
3. Zero the transducer again.
4. Replace the transducer.
Zeroing of IBP channel(s) fails. 1. Ensure that the channels are open to air.
2. Perform zero calibration again.
Do not sway the IBP transducer and
tubing during zero calibration.
For more information, see 14.3.3Zeroing
the IBP transducer
.
3. If zero calibration still fails, replace the transducer.