2. Wedge the flotation catheter into the pulmonary artery by observing the PA waveform changes on the
screen, referring to the following figure.
3. Select
Start.
4. Inflate the balloon and pay attention to PA waveform changes on the screen when the prompt message
Ready For Balloon Deflation appears.
5. Deflate the balloon when the prompt message
Ready For Balloon Deflation appears. If the PA waveform
is stable, but the monitor is still not showing the prompt message
Ready For Balloon Deflation, select
Freeze to freeze the waveform, and deflate the balloon.
6. Select
Accept to save the PAWP value.
7. If you need to start a new measurement, repeat step 3 to step 6.
If the measurement fails or you need to adjust the PAWP value, you can use the following buttons to adjust the
PAWP waveform and measurement.
Select the up or down arrow button to adjust the PAWP value.
Select the left or right arrow button to view the frozen waveforms of 40 seconds.
Select Accept to save the PAWP value.