14.3 Preparing for IBP Monitoring
14.3.1 IBP Equipment to Patient Connection
14.3.2 Measuring an Invasive Blood Pressure
To monitor IBP, follow this procedure:
1. Connect one end of the IBP cable to the IBP cable connector, and the other end to the IBP transducer.
2. Flush the IBP transducer system to exhaust all air from the tubing according to the manufacturer’s
instructions. Ensure that the system is free of air bubbles.
3. Connect the IBP transducer to the patient, making sure that the transducer is at the same horizontal level as
the heart.
4. Select the proper pressure label for currently measured pressure. For more information, see
14.6.2Changing the Pressure Label.
5. Zero the IBP transducer. For more information, see.
14.3.3Zeroing the IBP transducer. After a successful
zeroing, turn off the three-way valve to the air and turn on the three-way valve to the patient.
CAUTION
Make sure that all the transducers are zeroed correctly before making IBP measurements.
Make sure that all air bubbles have been flushed from the IBP transducer system before making IBP
measurements.
If measuring intracranial pressure (ICP) with a sitting patient, level the transducer with the top of the
patient’s ear. Incorrect leveling may give incorrect values (not applicable if measuring ICP with the
Codman ICP transducer).
(1) Pressure bag/Infusion pump (2) Monitor
(3) IBP cable (4) IBP transducer
(5) Three-way valve
mm Hg