4.2.6 Exergen TemporalScanner Thermometer Test
Tool required:
Water bath
Blackbody as specified in EN 12470-5-2003
Four-channel thermometer (model: 1529-R)
1. Put the probe of four-channel thermometer into water bath, and then put the
blackbody into water bath, set the water bath to 37.0. allow the water bath
sufficient time to equilibrate.
2. Set the Exergen TemporalScanner Themometer to EAR mode and ”.
3. Install a new probe cover. Measure and record the target temperature of blackbody
with the Exergem thermometer and four-channel thermometer.
4. Verify accuracy of temperature on monitor(subtract temperature value on
four-channel thermometer from temperature value on monitor) is within±0.1.
5. Repeat sections 1 to 4. For water bath set to 42.0.
6. Verify accuracy of temperature on monitor is within ±0.2.
Please contact our service personnel if the Temp test fails.
4.2.7 SmarTemp
TM
Module Test
Required Tool:
Thermostatic oil tank (HART 7102 recommended) or Water Bath and Precision
Thermometer for reference
Test procedure:
1. Set the temperature of the oil tank or water bath to 37°C and conduct the test
after the temperature stabilizes.
2. Set temperature unit to °C.
3. Select temperature parameter area to access [
Temp Setup] menu, and then set
temperature type to [
Monitor].
4. In Monitor Mode, remove the Temperature probe from the probe sheath, insert a
probe cover, and place the probe into the oil tank or water bath.
5. Wait till the Temp value displayed on the monitor stabilizes. Verify that the
displayed value is ±0.1°C of the oil bath thermostat setting or reference
thermometer.
Contact our Technical Support Department if the Temperature test fails.
NOTE
Due to the different application environment and the test object in vivo and
vitro conditions, there are deviations in the measurement result. The