Product Manual
4-2
4.2.1 Performance Test Frequencies
Check/Maintenance Item Frequency
Visual inspection 1. When first installed or reinstalled.
SpO
2
test
1. If the user suspects that the measurement is
incorrect.
2. Following any repairs or replacement of
relevant module.
3.
Every two years for SmarTemp
TM
module test
and SpO
2
test.
4.
Per year is recommended for NIBP and CO
2
tests.
NIBP test
Leakage test
Accuracy test
NIBP cuff
overpressure test
CO
2
test
(Accutorr 7 only)
Performance test
Calibration
TEMP test
Exergen
TemporalScanner
thermometer
SmarTemp
TM
module
4.2.2 Visual Inspection
Perform an overall inspection on the appearance of the equipment. The test is passed if the
equipment has no obvious signs of damage. Follow these guidelines when inspecting the
equipment:
Carefully inspect the case, display screen, buttons, and knob for obvious signs of
damage.
Inspect all external connections for loose connectors, bent pins or frayed cables.
Inspect all connectors on the equipment for loose connectors or bent pins.
Make sure that safety labels and data plates on the equipment are clearly legible.
4.2.3 SpO
2
test
Test Method 1
Tool required:
None
Test procedure:
1. Connect SpO
2
sensor for adult to the SpO
2
connector of the monitor. Set [Patient
Category
] to [Adult].
2. Place the SpO
2
sensor on your finger.
3. Verify the Pleth Wave and Pulse Rate are displayed on the screen.
4. Remove the SpO
2
sensor from your finger and make sure that the SpO
2
Sensor Off
alarm is triggered.