Panorama™ Operating Instructions 0070-10-0690-01 4 - 7
Arrhythmia and ST Analysis Lead Placement Procedures
Modified Chest Lead (MCL) Monitoring
The recommended lead placement for MCL monitoring is as follows.
FIGURE 4-9 MCL Monitoring with a
3-wire Lead Set (AHA)
FIGURE 4-10 MCL Monitoring with a
3-wire Lead Set (IEC)
Place RA (white) electrode under left clavicle,
mid-clavicular line within the rib cage frame.
Place LA (black) electrode on the right sternal
border, fourth intercostal in the rib cage
frame.
Place LL (red) electrode on the lower left
abdomen within the rib cage frame.
Place R (red) electrode under left clavicle,
mid-clavicular line within the rib cage frame.
Place L (yellow) electrode on the right sternal
border, fourth intercostal in the rib cage
frame.
Place F (green) electrode on the lower left
abdomen within the rib cage frame.
Select ECG Lead I for MCL
1
monitoring.
Lead I is the direct electrical line between
the RA (white) electrode and the LA (black)
electrode.
Select ECG Lead II for MCL
6
monitoring.
Lead II is the direct electrical line between
the RA (white) electrode and the LL (red)
electrode.
Select ECG Lead I for MCL
1
monitoring.
Lead I is the direct electrical line between the
R (red) electrode and the L (yellow)
electrode.
Select ECG Lead II for MCL
6
monitoring.
Lead II is the direct electrical line between
the L (red) electrode and the F (green)
electrode.