Lead Placement Procedures Arrhythmia and ST Analysis
4 - 6 0070-10-0690-01 Panorama™ Operating Instructions
Lead II Monitoring
The recommended lead placement for Lead II monitoring is as follows.
FIGURE 4-7 Lead II Monitoring (AHA) FIGURE 4-8 Lead II Monitoring (IEC)
Place RA (white) electrode under right
clavicle, mid-clavicular line within the rib
cage frame.
Place LA (black) electrode under left clavicle,
mid-clavicular line within the rib cage frame.
Place LL (red) electrode on the lower left
abdomen within the rib cage frame.
Place R (red) electrode under right
clavicle, mid-clavicular line within the rib
cage frame.
Place L (yellow) electrode under left
clavicle, mid-clavicular line within the rib
cage frame.
Place F (green) electrode on the lower
left abdomen within the rib cage frame.
Select the ECG Lead II at the bedside
monitor. Lead II is the direct electrical line
between the RA (white) electrode and the
LL (red) electrode.
Select the ECG Lead II at the bedside
monitor. Lead II is the direct electrical
line between the R (red) electrode and
the F (green) electrode.