User Manual
Table Of Contents
- 08-0308 - Manual Cover.pdf
- 0070-10-0690-01 Panorama Operations Maual with Telepack 2.4 - CH16 REVISED2.pdf
- Foreword
- Warnings, Precautions and Notes
- Indications for Use
- Unpacking Information
- Symbols and Descriptions
- Technical Support
- 1.0 System Overview
- 1.1 Components of the Panorama System
- 1.2 Licensing Key
- 1.3 Supported Devices
- 1.3.1 Panorama Central Station
- 1.3.2 Panorama ViewStation and WorkStation
- 1.3.3 Panorama Telemetry Server
- 1.3.4 Panorama Wireless Transceiver
- 1.3.5 Panorama Gateway
- 1.3.6 Panorama Web Viewer
- 1.3.7 Passport 2® Bedside Monitor
- 1.3.8 Spectrum™ Bedside Monitor
- 1.3.9 Spectrum OR™ Bedside Monitor
- 1.3.10 Panorama Telepack
- 1.3.11 Network Printer
- 1.4 Troubleshooting
- 2.0 Getting Started
- 3.0 Alarm Behavior
- 4.0 Arrhythmia and ST Analysis
- 5.0 Patient Setup Functions
- 5.1 Demographics Tab
- 5.2 Transfer Tab
- 5.3 Discharge Tab
- 5.4 Display Tab
- 5.5 Standby Tab
- 5.6 Wave Gain Tab (Patient)
- 5.7 Print Setup Tab (Patient)
- 6.0 Patient Alarm Functions
- 7.0 Views
- 7.1 Bedside Tab
- 7.2 Trends Tab
- 7.3 Events Tab
- 7.4 Disclosure Tab
- 7.4.1 Accessing the Disclosure Tab
- 7.4.2 Disclosure Tab (Waveforms View)
- 7.4.2.1 Accessing the Waveforms View
- 7.4.2.2 Components of the Waveforms View
- 7.4.2.3 Navigating in the Waveforms View of the Disclosure Tab
- 7.4.2.4 ST Button
- 7.4.2.5 Comments Dialog Box (Optional)
- 7.4.2.6 Sidebar Buttons
- 7.4.2.7 Printing a Full Disclosure Zoom In Report from the Waveforms View
- 7.4.2.8 Troubleshooting
- 7.5 ST Tab
- 8.0 Report Functions
- 8.1 Patient Reports Tab
- 8.2 System Reports Tab
- 8.3 Print Status Tab
- 8.4 Additional Reports
- 9.0 System Setup Functions
- 9.1 Parameter Color Tab
- 9.2 Touch Screen Tab
- 9.3 Installation Setup Tab
- 9.4 Care Group Tab
- 9.5 Volume Tab
- 9.6 System Alarms Tab
- 9.7 Passwords Tab
- 9.8 Equipment Setup Tab
- 9.9 More Tab
- 9.10 Previous Tab
- 9.11 Date/Time Tab
- 9.12 Print Setup Tab (System)
- 9.13 Unit Priorities Tab
- 9.14 Unit Choices Tab
- 9.15 Wave Gain Tab (System)
- 9.16 Wireless Tab
- 10.0 Telepack (Optional)
- 11.0 User Maintenance
- 12.0 Parameter Specifications
- 12.1 Parameter Specifications
- 12.1.1 Heart Rate: HR
- 12.1.2 Respiration: Resp
- 12.1.3 NIBP
- 12.1.4 SpO2
- 12.1.5 Nonin SpO2
- 12.1.6 IBP: mmHg
- 12.1.7 CO2
- 12.1.8 O2
- 12.1.9 INSP/ET O2
- 12.1.10 N2O
- 12.1.11 Anesthetic Agents: Iso, Enf, Des, Sev, Hal, Agent
- 12.1.12 Premature Ventricular Contraction (PVC)
- 12.1.13 Temp: T1,T2, Delta T, and T Blood
- 12.1.14 ST
- 12.1.15 Cardiac Output
- 12.1.16 Cardiac Index
- 12.1.17 Pulmonary Artery Wedge Pressure (PAWP)
- 12.1.18 Battery Percentage
- 12.2 Invalid Data Display
- 12.1 Parameter Specifications
- 13.0 Alarms and Events
- 14.0 Network Tab (Optional)
- 14.1 Network Tab
- 14.1.1 Accessing the Network Tab
- 14.1.2 Sorting the Contents of the Network Tab List Box
- 14.1.3 Assigning Patient Tiles in the Network Tab
- 14.1.4 Removing a Patient Tile Assignment in the Network Tab
- 14.1.5 Removing All Patient Tile Assignments in the Network Tab
- 14.1.6 Refreshing the Network Tab List Box
- 14.1.7 Troubleshooting
- 14.1 Network Tab
- 15.0 Paging Tab (Optional)
- 15.1 Paging Password Tab
- 15.2 Paging Tab
- 15.2.1 Accessing the Paging Tab
- 15.2.2 Adding a Pager
- 15.2.3 Assigning a Patient Tile
- 15.2.4 Adding All Patient Tiles to a Pager
- 15.2.5 Removing a Patient Tile Assignment
- 15.2.6 Removing All Patient Tiles from a Pager
- 15.2.7 Editing a Pager
- 15.2.8 Removing a Pager
- 15.2.9 Resending a Failed Page
- 15.2.10 Generating a Paging Report
- 15.2.11 Troubleshooting
- 17.0 Glossary
Physiological Alarms Alarm Behavior
3 - 4 0070-10-0690-01 Panorama™ Operating Instructions
Once a latched alarm condition is acknowledged, the alarm indicators discontinue until
another alarm condition occurs. Acknowledging the alarm at the central station will not
discontinue the alarm at the bedside monitor, and acknowledging the alarm at the bedside
monitor will not discontinue the alarm at the central station, ViewStation, or WorkStation.
Apnea Latching
The conditions for apnea latching follow standard latching procedures, however apnea
latching is associated with an apnea alarm condition. This option can be enabled/disabled
in the System Alarms (System Alarm Options) tab. By default, Apnea Latching is enabled.
NOTE: Respiratory apnea alarm condition events are detected by
the bedside monitor.
NOTE: Acknowledging an apnea latched alarm at the central
station patient tile also acknowledges the alarm at the
associated ViewStation and WorkStation patient tile.
NOTE: Acknowledging an apnea latched alarm at the bedside
monitor does not acknowledge the alarm at the host central
station, ViewStation, or WorkStation patient tile.
Alarm Delay
Alarm Delay is the configurable time period required before an alarm is triggered. Alarm
Delay will not effect the Apnea Alarm, ST Alarms, the non-lethal arrhythmia alarms, or the
lethal arrhythmia alarms. When permitted, this option can be enabled/disabled in the
System Alarms (System Alarm Options) tab. By default, an Alarm Delay time is not
specified.
ST Alarm Delay
ST alarm delay postpones the trigger of the ST alarm for a configured time period. When
permitted, this option can be configured in the System Alarms (System Alarm Options)
tab.
3.2.3 Physiological Alarm Responses
The alarm responses associated with a physiological alarm are as follows.
3.2.3.1 Alarm Priorities
The Panorama Central Station supports three alarm priorities. When alarm priority is
turned on, the choices available are Alarm Priority of 1, 2 or 3. Alarms may be turned off by
using the Priority Off setting. Alarm priorities associated with lethal alarm parameters cannot
be changed from their default value of Priority 1. However, all other parameters can be
modified. The alarm priority used to configure a parameter defines the alarm severity.
When a parameter is assigned a Priority Off, there are no visual or audio alarm indicators
associated with the alarm event, however Print on Alarm and Save To Event options
are still available.
Alarm priorities are configured in the patient Alarm Responses and the System Alarms
(System Alarm Responses) tabs.