2.6 Troubleshooting
This section lists some of the potential issues that may occur when admitting a patient to the
Panorama Central Station.
MESSAGE/ISSUE * REASON SOLUTION
Cannot admit patient/patient
data not displayed at central
station.
Previous patient not discharged
from system and/or monitoring
device.
Discharge patient from system.
Monitoring device not
assigned to Panorama
Equipment List.
Assign device to Panorama
Equipment List.
Panorama not installed as an
option (for
Passport 2
®
/Spectrum
only).
Access the Installation Menu,
Options submenu at the
bedside device. For additional
information, refer to the
Passport 2, Spectrum, or
Spectrum OR
Operating
Instructions.
Network setting on monitoring
device is incorrect.
Access the Installation Menu,
System Information submenu at
the bedside device. For
additional information, refer to
Passport 2, Spectrum, or
Spectrum OR
Operating
Instructions.
Telepack batteries may need to
be replaced.
Test batteries, and if necessary,
replace.
Telepack batteries improperly
installed.
Reinstall batteries.
* Messages are shown in all bold text.