2 - 20 0070-10-0690-01 Panorama™ Operating Instructions
2.5 Quick Reference Guide
The quick reference guide provides abbreviated procedural information for the operation of
the Panorama Central Station.
For additional information, refer to the Panorama Quick Reference Guide
(P/N 0002-08-7538).
Quick reference procedures include:
Auto-Admitting a Patient
•Standby
•Discharge
Display
Patient Alarms (Limits and Responses)
Viewing List Trends
Viewing Events
Viewing Disclosure Data
NOTE: Demographic information is bi-directionally shared between
the bedside monitor and the Panorama Central Station.
2.5.1 Auto-Admitting a Patient
Auto-admit allows for the immediate monitoring of a patient at the Panorama Central Station.
To auto-admit a patient:
1. Verify that the monitoring device is in the Equipment List. For information on how to
access the Equipment List, refer to the “Equipment Setup Tab” on page 9-33.
2. Connect/attach the monitoring device to the patient.
3. Turn on the monitoring device. The monitoring device begins monitoring the patient. The
parameters that are monitored are dependent on the monitoring device.
4. Verify that the patient is being monitored at the central station.
NOTE: Auto-admit uses the system default settings for monitoring
the patient.
2.5.2 Standby
Standby mode suspends the monitoring of a patient at the Panorama Central Station while
retaining the patient’s historical data and setup priorities. It is important that a patient in this
mode be alternatively monitored. The Standby tab is also available at the WorkStation.
To activate Standby mode at the central station or the WorkStation:
1. Select the Patient Setup menu button.
2. Choose a patient tile by selecting its digital data or waveform area.
3. Select the Standby tab.
•The Panorama Standby Only button places the patient in Standby at the
Panorama Central Station.