User Manual

Table Of Contents
Panorama™ Operating Instructions 0070-10-07XX-02 16 - 27
Performance Specifications Nonin SpO
2
Module Specifications
16.11 Nonin SpO
2
Module Specifications
16.11.1 Environmental Characteristics
The environmental characteristics for the Nonin SpO
2
module device are:
16.11.2 Measurement Specifications
The measurement specifications for the Nonin SpO
2
module are:
TABLE 16-12
SPECIFICATION DESCRIPTION
Maximum Size of Telepack
with Nonin SpO
2
module
Height: 6.25” (158.75 mm)
Width: 3.55” (90.17 mm)
Depth: 1.75” (44.45 mm)
Maximum Weight
(Nonin SpO
2
module)
4.5 oz. excluding cable and sensor
Temperature Operating: + 5
°C to + 40 °C
Storage: - 20
°C to + 50 °C
Humidity Operating:10% to 90% maximum, non-condensing
Storage: 10% to 95% maximum, non-condensing
SPECIFICATION DESCRIPTION
Oxygen Saturation Range 0 to 100%
Pulse Rate Range 18 to 300 pulses per minute
Measurement wavelength red 660 nanometers @3mW nominal
NOTE: Information about wavelength range
can be especially useful to clinicians.
Measurement wavelength infrared 910 nanometers @3mW nominal
NOTE: Information about wavelength range
can be especially useful to clinicians.
SPO2 ACCURACY
SpO
2
(70-100%) (±1 standard deviation)
No Motion reusable finger clip ± 2 digits
disposable ± 3 digits
low perfusion ± 2 digits
Motion reusable finger clip ± 2 digits
low perfusion ± 2 digits
PULSE RATE ACCURACY
no motion adults (18-300 bpm): ± 3 digits
motion adults (40-240 bpm): ± 5 digits
low perfusion adults (40-240 bpm): ± 3 digits
DATA UPDATE RATE
SpO