Panorama™ Operating Instructions 0070-10-07XX-02 16 - 23
Performance Specifications Telepack Specifications
Voltage dips,
short
interruptions and
voltage
variations on
power supply
input lines IEC
61000-4-11
<5% U
Τ
(>95%
dip in U
T
) for
0,5 cycle
40% U
T
(60%
dip in U
T
) for
5 cycles
70% U
T
(30%
dip in U
T
) for
25 cycles
< 5% U
T
(>95%
dip in U
T
) for 5
sec
N/A * Mains power quality should be typical
of a commercial or hospital
environment. If the user requires
continued operation during power
mains interruptions, it is recommended
that the device be powered from an
uninterruptible power supply or a
battery.
Power frequency
(50/60 Hz)
magnetic field
IEC 61000-4-8
3 A/m 3 A/m Power frequency magnetic fields
should be at levels characteristic of a
typical location in a typical commercial
or hospital environment.
TABLE 16-9 (Continued)
GUIDANCE AND DATASCOPE CORP. DECLARATION - ELECTROMAGNETIC IMMUNITY
It is important that the Telepack is used in the electromagnetic environment specified in the following
table.
IMMUNITY
TEST
IEC 60601
TEST LEVEL
COMPLIANCE
LEVEL
ELECTROMAGNETIC
ENVIRONMENT - GUIDANCE
U
T
is the a.c. mains voltage prior to application of the test level.
* Telepack operates only from the internal battery