User Manual

Table Of Contents
Panorama™ Operating Instructions 0070-10-07XX-02 16 - 21
Performance Specifications Telepack Specifications
16.10.3 Environmental Characteristics
The environmental characteristics of the Telepack device are as follows:
16.10.4 Electromagnetic Compatibility
The following section describes the electromagnetic compatibility of the Telepack device.
NOTE: The Telepack needs special precautions regarding Electro
Magnetic Compatibility (EMC) and needs to be installed and
put into service according to the EMC information provided
in this section.
NOTE: Portable and mobile RF communications equipment can
affect the Telepack. See Table 16-8 on page 16-22,
Table 16-9 on page 16-22, Table 16-10 on page 16-24 and
Table 16-11 on page 16-25.
TABLE 16-7
SPECIFICATION DESCRIPTION
Maximum Size (Telepack) Height: 5.10” (129.54 mm)
Width: 2.95” (74.93 mm)
Depth: 1.23” (31.24 mm)
Operating Environment Temperature: + 5
°C to + 40 °C (+41 °F to +104 °F)
Altitude:
1060 hPa to 700 hPa
-1250 feet to 9889 feet ASL
-380 m to 3014 m
795 mmHg to 525 mmHg
Operating and Storage
Humidity
Storage: 10% to 90% maximum, non-condensing
Operating: 15% to 90% maximum, non-condensing
Storage Temperature --20 °C to +60 °C (-4 °F to +140 °F)
Storage Altitude 1,060 hPa to 700 hPa
-1250 feet to 9,889 feet ASL
-380 m to 3,014 m
795 mmHg to 525 mmHg