16.10 Telepack Specifications
The following are the specifications for the Telepack:
16.10.1 Safety Designations
16.10.2 Panorama Server Analysis
The bedside monitor collects and analyses patient information and then transfers its analysis
to the Panorama Central Station through the Panorama Telemetry Server; therefore, no
calculations are performed by the Panorama Telemetry Server.
TABLE 16-5
SPECIFICATION DESCRIPTION
Acquired Leads ECG 3-Lead: I, II, III (one vector at a time)
ECG 5-Lead: Three vectors (I, II, III, aVR, aVF, V(n))
Battery Runtime/
Battery Type
24 hrs minimum for Telepack-2.4 only,
16 hrs minimum for Telepack-2.4 with Nonin SpO
2
module,
40 hrs minimum for Telepack-608 only,
24 hrs minimum for Telepack-608 with Nonin SpO
2
module
using (2) AA alkaline cells (P/N 0146-00-0077-XX) with ECG at
a HR of 60 bpm common to the Panorama Central Station
Battery Shelf Life Indicated on battery casing
Lead Current Max lead current: 25nA
TABLE 16-6
DESIGNATION DESCRIPTION
Degree of protection against
electric shock
ECG-Type CF defibrillation protected.
Supply Connection 3 VDC Internal Battery
Mode of Operation Continuous
Protection Against Hazards of
Explosion
Not Protected (Ordinary)
Degree of Electrical Connection
between Equipment and Patient
Equipment designed for direct electrical connection to the patient
Degree of Mobility Transportable, Intra Hospital
Protection against ingress of liquid (please refer to your device):
Protected against splashing
water
IPX4 (with or without SpO
2
module attached)