User Manual

Table Of Contents
Panorama™ Operating Instructions 0070-10-07XX-02 16 - 9
Performance Specifications Panorama Central Station and Panorama Telemetry Server
16.3.6 Storage Temperature
-20 to +60 º C
16.3.7 Storage Humidity
5% to 80% Relative Humidity Maximum, non-condensing, maximum wet bulb 35 º C
16.3.8 Storage Altitude
1060 to 700 hPa (-1250 to 9889 feet ASL) (-380 - 3014 m) (795 to 525 mm Hg)
16.3.9 Safety
The Central Station and Server comply with the following standards:
1. UL 60950-1 Safety of Information Technology Equipment
2. CSA-C22.2 60950-1-03 Safety of Information Technology Equipment
3. IEC 60601-1-8 Electrically-generated alarm signals
4. IEC 60601-1-2 Electromagnetic compatibility
5. ANSI/AAMI EC 13 Cardiac Monitors, heart rate meters, and alarms
6. ISTA Test Procedure 1A Shipping Integrity Test
SAFETY DESIGNATIONS
(SEE SECTION 16.3.10 (TABLE 16-1, TABLE 16-2, TABLE 16-3, TABLE 16-4)
Mode of Operation Continuous
Protection Against Hazards of Explosion Non Protected (ordinary)
Protection Against Ingress of Liquids IPX0 (ordinary, non-protected)
Degree of Mobility Stationary, Intra Hospital