Panorama™ Operating Instructions 0070-10-07XX-02 16 - 5
Performance Specifications ECG Derived Heart Rate Meter Performance Requirements
User Selectable ST
16.2.3 Arrhythmia Analysis
Arrhythmia analysis identifies ventricular arrhythmia.
The following arrhythmia calls will be displayed:
16.2.4 NIBP Performance Characteristics
Systolic Pressure Measurement
16.2.4.1 Diastolic Pressure Measurement
Measurement Points: 40, 60 and 80 ms after J point (heart rate independent)
Or 60/80 Heart Rate Dependent
Default ISO Point: Located between the P and Q waves.
User adjustable ISO Point: User adjustable from “R peak” - 10 ms to “R peak” -
200 ms in 8 ms increments.
Default J Point: The end of the QRS complex.
User adjustable J Point: User adjustable from “R peak” - 10 ms to “R peak” -
200 ms in 8 ms increments.
Numerical ST Update Rate: 3/5 Lead Wire: every 10 seconds
View 12 card: every 2 seconds
Asystole Pause
Bigeminy Run
Bradycardia (Brady) Trigeminy
Couplet Ventricular Fibrillation (V-Fib)
Irregular Heart Rate Ventricular Rhythm
PVCs per minute Ventricular Tachycardia (V-Tach)
Range: 55 to 235 mmHg in Adult mode
55 to 160 mmHg in Pediatric mode
45 to 120 mmHg in Neonatal mode
Resolution: 1 mmHg
Range: 30 to 200 mmHg in Adult mode
30 to 150 mmHg in Pediatric mode
20 to 100 mmHg in Neonatal mode
Resolution: 1 mmHg