Technical Alarms and Events Alarms and Events
13 - 12 0070-10-0690-01 Panorama™ Operating Instructions
Low Battery
5
The monitoring equipment battery is low.
NOTE: If the Low Battery message
appears and the monitoring
device is a Telepack, the Battery
Percentage digital data tile will
flash yellow until the issue is
resolved.
Cooling Fan Failure The bedside monitor cooling fan is not
operational and the monitor is running on AC
power.
Admit
2,3
A patient is admitted to the Panorama Central
Station.
Discharge The patient was discharged from the Panorama
Central Station.
Arrhythmia On
1,2,3
The Arrhythmia option is enabled on a patient’s
monitoring equipment.
Arrhythmia Off
1,2,3
The Arrhythmia option is disabled on a patient’s
monitoring equipment.
ST On
1,2,3
The ST option is enabled on a patient’s
monitoring equipment.
ST Off
1,2,3
The ST option is disabled on a patient’s
monitoring equipment.
ST Relearn
2,3
The ST algorithm completes a learning phase.
Arrhythmia Relearn
2,3
The arrhythmia algorithm completes a learning
phase.
Central User Marked Event
2,3
The user has marked an event at the central
station.
Bedside User Marked Event
2,3
The user has marked an event at the bedside
monitor.
Calipers Marked Event
2,3
The user has marked a caliper event at the
Panorama Central Station
Central Transfer
2,3
A central-to-central transfer has been successfully
completed at the central station.
Room Transfer
2,3
A room-to-room transfer has been successfully
completed at the central station.
HEART RATE TECHNICAL EVENTS
Check Lead Connections The bedside monitor detects a lead off condition.
RA Lead Off
4,5
The RA lead is not connected to the patient, ECG
lead I or II is selected while using a 3-wire lead
set, 5-wire lead set or a View 12
card.
LA Lead Off
4,5
The LA lead is not connected to the patient, ECG
lead I or III is selected while using a 3-wire lead
set, 5-wire lead set, or a View 12
card.
Technical Events (Continued)
MESSAGE REASON
1 Event cannot be deleted from Event database
2 Event will not elicit technical event sound
3 Event will not display in Patient Status line
4 Event will always elicit the technical event sound regardless of the Technical Event sound setting
5 Event will post a Page request when the central station has the paging option enabled and the alarm is
configured to send a page