User Manual

Table Of Contents
Panorama™ Operating Instructions 0070-10-0690-01 13 - 7
Alarms and Events Physiological Alarms and Events
13.1.2 Physiological Alarm Event Messages
This section lists the alarm messages that may appear at the Panorama Central Station when
an Physiological alarm event occurs.
PARAMETER
PANORAMA CENTRAL
STATION MESSAGES REASON
Heart Rate (HR) High Heart Rate Alarm
Low Heart Rate Alarm
Current value is either equal to or higher/lower
than the set alarm limits.
Respiration (ECG
Thoracic
Impedance, CO
2,
and Gas Module)
High Respiration Rate
Alarm
Low Respiration Rate
Alarm
Current value is either equal to or higher/lower
than the set alarm limits.
SpO
2
(Masimo
SET
®
, Nellcor
®
,
Nonin)
High SpO
2
Alarm
Low SpO
2
Alarm
Current value is either equal to or higher/lower
than the set alarm limits.
IBPx (Sys) Labels
include: IBP1,
IBP2, IBP3, IBP4,
ART, UA, LV, PA,
CVP, ICP, LA, RA
High IBPx Systolic Alarm
Low IBPx Systolic Alarm
Current value is either equal to or higher/lower
than the set alarm limits.
IBPx (Dia) Labels
include: IBP1,
IBP2, IBP3, IBP4,
ART, UA, LV, PA,
CVP, ICP, LA, RA
High IBPx Diastolic Alarm
Low IBPx Diastolic Alarm
Current value is either equal to or higher/lower
than the set alarm limits.
IBPx (Mean)
Labels include:
IBP1, IBP2, IBP3,
IBP4, ART, UA,
LV, PA, CVP, ICP,
LA, RA
High IBPx Mean Alarm
Low IBPx Mean Alarm
Current value is either equal to or higher/lower
than the set alarm limits.
NOTE: x = 1, 2, 3 or 4 (IBP3 and 4 are only available if the
Spectrum/Spectrum OR External Parameter Module
(EPM) is installed.)
INSP CO
2
(mmHg, %, kPa)
High CO
2
Inspired Alarm Current value is either equal to or higher than
the set alarm limits.
ET CO
2
(mmHg,%, kPa)
High CO
2
ET Alarm
Low CO
2
ET Alarm
Current value is either equal to or higher/lower
than the set alarm limits.
INSP O
2
O
2
Inspired Alarm There has been a violation of the O
2
Inspired
alarm.
ET O
2
O
2
ET Alarm There has been a violation of the O
2
ET alarm.
INSP N
2
ON
2
O Inspired Alarm There has been a violation of the N
2
O Inspired
alarm.
ET N
2
ON
2
O ET Alarm There has been a violation of the N
2
O ET alarm.
Agent Agent Inspired Alarm There has been a violation of the Agent Inspired
alarm.
Agent ET Alarm There has been a violation of the Agent ET
alarm.
INSP Iso Agent Iso Inspired Alarm There has been a violation of the Iso Inspired
alarm.
ET Iso Agent Iso ET Alarm There has been a violation of the Iso ET alarm.
INSP Enf Agent Enf Inspired Alarm There has been a violation of the Enf Inspired
alarm.