Physiological Alarms and Events Alarms and Events
13 - 6 0070-10-0690-01 Panorama™ Operating Instructions
T Blood
Units: F°
2 or 3
Factory Default: 3
Low Limit N/A Off, 80-100°
Factory Default: Off
Off, 80-100°
Factory Default: Off
Spectrum
/Spectrum OR
High Limit N/A Off, 95-110°
Factory Default: Off
Off, 95-110°
Factory Default: Off
T Blood
Units: C°
2 or 3
Factory Default: 3
Low Limit N/A Off, 26-38°
Factory Default: Off
Off, 26-38°
Factory Default: Off
Spectrum/Spectrum OR
High Limit N/A Off, 35-43°
Factory Default: Off
Off, 35-43°
Factory Default: Off
NOTE: Only available if Spectrum/Spectrum OR External Parameter Module (EPM) is installed and sensor is
attached
ST Single
Units: mm
1, 2 or 3
Factory Default: 2
High Limit N/A Off,
0.50 to10.0
Factory Default: Off
Off,
0.50 to 10.0
Factory Default: Off
Passport 2
Spectrum/Spectrum OR
Telepack
Low Limit N/A Off,
(-0.5) to
(-10.0)
Factory Default: Off
Off,
(-0.5) to
(-10.0)
Factory Default: Off
ST Dual
Units: mm
1, 2 or 3
Factory Default: 2
High Limit N/A Off,
0.50-10.0
Factory Default: Off
Off,
0.50-10.0
Factory Default: Off
Spectrum/Spectrum OR
Telepack
Low Limit N/A Off,
(-0.5) - (-10.0)
Factory Default: Off
Off,
(-0.5) - (-10.0)
Factory Default: Off
PVC
Units:
PVCs/min
2 or 3
Factory Default: 3
High Limit N/A Off, 1-30
Factory Default: Off
Off, 1-30
Factory Default: Off
Passport 2
Spectrum/Spectrum OR
Telepack
Low Limit N/A N/A N/A
Apnea Delay
Units: Seconds
2 or 3
Factory Default: 3
High Limit 10-20
Factory Default: 15
10-20
Factory Default: 15
10-60
Factory Default: 60
Passport 2
Spectrum/Spectrum OR
Low Limit
ALARM
ALARM
PRIORITY ALARM LIMIT NEONATE PEDIATRIC ADULT DEVICES
N/A - Not Applicable