Physiological Alarms and Events Alarms and Events
13 - 4 0070-10-0690-01 Panorama™ Operating Instructions
IBPx Dia.
Units: mmHg*
2 or 3
Factory Default: 3
High Limit Off, 0-70
Factory Default: Off
Off, 0-100
Factory Default: Off
Off, 0-140
Factory Default: Off
Passport 2
®
Spectrum
/Spectrum OR
Low Limit Off, 0-50
Factory Default: Off
Off, 0-100
Factory Default: Off
Off, 0-120
Factory Default: Off
IBPx Mean
Units: mmHg*
2 or 3
Factory Default: 3
High Limit Off, 5-100
Factory Default: Off
Off, 5-100
Factory Default: Off
Off, 5-150
Factory Default: Off
Passport 2
Spectrum/Spectrum OR
Low Limit Off, 2-50
Factory Default: Off
Off, 2-50
Factory Default: Off
Off, 2-100
Factory Default: Off
All IBPx alarm limits apply to IBP1, IBP2, IBP3, IBP4, ART, UA, LV, PA, CVP, ICP, LA, and RA.
The Panorama Central Station is capable of monitoring up to four IBP measurements.
NOTE: x = 1, 2, 3 or 4 (IBP3 and 4 only available if Spectrum/Spectrum OR External Parameter Module
(EPM) is installed)
NIBP Sys.
Units: mmHg
2 or 3
Factory Default: 3
High Limit Off, 40-180
Factory Default: Off
Off, 40-180
Factory Default: Off
Off, 70-240
Factory Default: Off
Passport 2
Spectrum
/Spectrum OR
Low Limit Off, 15-130
Factory Default: Off
Off, 15-130
Factory Default: Off
Off, 15-150
Factory Default: Off
NIBP Dia.
Units: mmHg
2 or 3
Factory Default: 3
High Limit Off, 50-100
Factory Default: Off
Off, 50-100
Factory Default: Off
Off, 40-130
Factory Default: Off
Passport 2
Spectrum/Spectrum OR
Low Limit Off, 10-50
Factory Default: Off
Off, 10-50
Factory Default: Off
Off, 30-120
Factory Default: Off
NIBP Mean
Units: mmHg
2 or 3
Factory Default: 3
High Limit Off, 40-160
Factory Default: Off
Off, 50-180
Factory Default: Off
Off, 60-200
Factory Default: Off
Passport 2
Spectrum/Spectrum OR
Low Limit Off, 10-70
Factory Default: Off
Off, 10-100
Factory Default: Off
Off, 40-140
Factory Default: Off
INSP CO
2
(mmHg) 2 or 3
Factory Default: 3
High Limit Off, 5-30
Factory Default: Off
Off, 5-30
Factory Default: Off
Off, 5-30
Factory Default: Off
Passport 2
Spectrum/Spectrum OR
Low Limit N/A N/A N/A N/A
INSP CO
2
(%) 2 or 3
Factory Default: 3
High Limit Off, 1-4
Factory Default: Off
Off, 1-4
Factory Default: Off
Off, 1-4
Factory Default: Off
Passport 2
Spectrum/Spectrum OR
Low Limit N/A N/A N/A N/A
INSP CO
2
(kPa) 2 or 3
Factory Default: 3
High Limit Off, 1-4
Factory Default: Off
Off, 1-4
Factory Default: Off
Off, 1-4
Factory Default: Off
Passport 2
Spectrum/Spectrum OR
Low Limit N/A N/A N/A N/A
ALARM
ALARM
PRIORITY ALARM LIMIT NEONATE PEDIATRIC ADULT DEVICES
N/A - Not Applicable