User Manual

Table Of Contents
Panorama™ Operating Instructions 0070-10-0690-01 12 - 11
Parameter Specifications Invalid Data Display
12.2 Invalid Data Display
The Panorama Central Station uses dashes (---), Xs (XX) or a straight line (waveform) to show
invalid data.
Dashes (---) Digital Data
Digital data is replaced with dashes (---) for the following reasons:
Data is above the maximum value
Data is below the minimum value
Parameter is not available
Lead or a sensor is off
Sensor is not zeroed
Parameter is zeroing
Measurement has not been taken
Measurement is being taken
Measurement has timed out
Measurement is being retried
Parameter is turned off
•View 12
card is relearning
PVC/min counter reset
Gas module or the source agent is occluded
Gas module or the source agent is purging
Gas module is turned off
ST license is not available
Learning ST Values
Xs (XX) Digital Data
Digital data is replaced with Xs (XX) for the following reasons:
IBP parameter is not zeroed
NIBP parameter failed to obtain a reading