User Manual

Table Of Contents
Parameter Specifications Parameter Specifications
12 - 8 0070-10-0690-01 Panorama™ Operating Instructions
The Panorama Central Station supports Iso, Enf, Sev, Hal and Des waveforms.
The Panorama Central Station displays a Generic Agent waveform when the Gas Module
cannot identify a specific agent.
12.1.12 Premature Ventricular Contraction (PVC)
The Panorama Central Station supports PVC data.
12.1.13 Temp: T1,T2, Delta T, and T Blood
The Panorama Central Station supports four temperature measurements.
T1 and T2
Delta T Values
Unit of Measure: PVC/min
Resolution: 1 PVC/min
PATIENT SIZE LEADS MINIMUM VALUE MAXIMUM VALUE
Neonate 3 lead, 5 lead, or
12 lead
Not Applicable Not Applicable
Pediatric 3 lead, 5 lead, or
12 lead
0 PVC/min Less than the Current HR value
(PVCs/min)
Adult 3 lead, 5 lead, or
12 lead
0 PVC/min Less than the Current HR value
(PVCs/min)
Units of Measure: F° and C°
Resolution: 0.1 F° and 0.1 C°
PATIENT SIZE MINIMUM VALUE (F°) MAXIMUM VALUE (F°)
Neonate 59 113
Pediatric 59 113
Adult 59 113
PATIENT SIZE MINIMUM VALUE (C°) MAXIMUM VALUE (C°)
Neonate 15 45
Pediatric 15 45
Adult 15 45
PATIENT SIZE MINIMUM VALUE (F°) MAXIMUM VALUE (F°)
Neonate 0 9.9
Pediatric 0 9.9
Adult 0 9.9
PATIENT SIZE MINIMUM VALUE (C°) MAXIMUM VALUE (C°)
Neonate 0 5.5