User Manual

Table Of Contents
Parameter Specifications Parameter Specifications
12 - 4 0070-10-0690-01 Panorama™ Operating Instructions
NIBP (Sys)
NIBP (Dia)
NIBP Mean
NIBP (ET)
12.1.4 SpO
2
The SpO
2
parameter includes measurements from Masimo SET
®
and Nellcor
®
SpO
2
sensors.
The Panorama Central Station supports SpO
2
waveforms.
PATIENT SIZE MINIMUM VALUE (mmHg) MAXIMUM VALUE (mmHg)
Neonate 45 120
Pediatric 55 160
Adult 55 235
PATIENT SIZE MINIMUM VALUE (mmHg) MAXIMUM VALUE (mmHg)
Neonate 20 100
Pediatric 30 150
Adult 30 200
PATIENT SIZE MINIMUM VALUE MAXIMUM VALUE
Neonate mmHg greater than the NIBP (Dia) value mmHg less than the NIBP (Sys) value
Pediatric mmHg greater than the NIBP (Dia) value mmHg less than the NIBP (Sys) value
Adult mmHg greater than the NIBP (Dia) value mmHg less than the NIBP (Sys) value
PATIENT SIZE MINIMUM VALUE (MIN.) MAXIMUM VALUE (MINUTES)
Neonate 0 999
Pediatric 0 999
Adult 0 999
Unit of Measure: %
Resolution: 1%
PATIENT SIZE MINIMUM VALUE (%) MAXIMUM VALUE (%)
Neonate --- 100
Pediatric --- 100
Adult --- 100