Parameter Specifications Parameter Specifications
12 - 2 0070-10-0690-01 Panorama™ Operating Instructions
12.1 Parameter Specifications
The parameters displayed on the Panorama Central Station are dependent on the type of
equipment connected to the Panorama system.
12.1.1 Heart Rate: HR
The Heart Rate parameter obtains data from ECG and IBP sources.
Heart Rate Source: ECG
The Panorama Central Station displays ECG waveforms when using 3-wire, 5-wire, and 12
lead cable. The ECG leads available at the Panorama Central Station are dependent on the
lead cable connected to the patient.
Heart Rate Source: IBP
12.1.2 Respiration: Resp
The Respiration parameter obtains data from ECG, CO
2
or Gas Module sources.
Unit of Measure: Beats Per Minute (bpm)
Resolution: 1 bpm
PATIENT SIZE MINIMUM VALUE (bpm) MAXIMUM VALUE (bpm)
Neonate 30 350
Pediatric 30 300
Adult 30 300
LEAD WAVEFORM LEADS AVAILABLE
3-wire lead set I, II, III
5-wire lead set I, II, III, aVR, aVL, aVF, V
View 12
card I, II, III, aVR, aVL, aVF, V1, V2, V3, V4, V5, V6
Unit of Measure: mV
PATIENT SIZE MINIMUM VALUE (bpm) MAXIMUM VALUE (bpm)
Neonate 30 350
Pediatric 30 300
Adult 30 300
Unit of Measure: Respirations Per Minute (rpm)
Resolution: 1 rpm