Panorama™ Operating Instructions 0070-10-0690-01 11 - 5
User Maintenance Telepack User Maintenance
Soap and water
Glutaraldehyde 2% (for example, Cidex)
Dilute chlorine bleach (sodium hypochlorite), 10% solution, freshly made in past 24 hours
Ethyl Alcohol 30% (for example, green soap tincture)
To clean the ECG lead sets, remove them from the Telepack, and follow these guidelines:
Use the cleaning solution to wipe and rinse the lead set.
The cleaning solution should be wiped away with a cloth moistened with distilled water.
A lint free cloth should be used to dry the lead set thoroughly.
WARNING: The Telepack should not be used adjacent to or stacked with
other equipment. If adjacent or stacked use is necessary, the
Telepack should be observed to verify normal operation in
the configuration in which it will be used.
Sterilization of the Lead Sets
CAUTION: Prior to sterilization, lead sets must be removed from the
Telepack. The Telepack must not be sterilized.
Use EtO to sterilize the leadsets. The sterilization should be performed by following the
hospital-approved procedure for EtO sterilization, such as those recommended by AAMI. The
leadsets remain effective up to 10 exposures to EtO sterilization cycles.
NOTE: Damage to the lead sets will result if steam or gamma
radiation is used for the sterilization process.
11.2.3 Cleaning the SpO
2
Module
In the event that the SpO
2
Module becomes soiled, a cleaning plug (P/N 0003-00-0055)
should be used to cover the serial connector on the SpO
2
Module before it is cleaned.
CAUTION: Datascope requires that the designated cleaning plug (P/N
0003-00-0055) be used when cleaning and disinfecting the
Nonin SpO
2
module. If fluids enter the device, it may affect
the functionality of the module or render it unusable.
Approved Cleaning Solutions
Please refer to the approved cleaning solutions described in ‘‘Cleaning and Disinfecting the
Telepack Device’’ on page 11-3.
Inserting the Cleaning Plug
1. Carefully cover the connector by aligning the cleaning plug with the lower portion of the
SpO
2
Module.