User Manual
Table Of Contents
- 08-0308 - Manual Cover.pdf
- 0070-10-0690-01 Panorama Operations Maual with Telepack 2.4 - CH16 REVISED2.pdf
- Foreword
- Warnings, Precautions and Notes
- Indications for Use
- Unpacking Information
- Symbols and Descriptions
- Technical Support
- 1.0 System Overview
- 1.1 Components of the Panorama System
- 1.2 Licensing Key
- 1.3 Supported Devices
- 1.3.1 Panorama Central Station
- 1.3.2 Panorama ViewStation and WorkStation
- 1.3.3 Panorama Telemetry Server
- 1.3.4 Panorama Wireless Transceiver
- 1.3.5 Panorama Gateway
- 1.3.6 Panorama Web Viewer
- 1.3.7 Passport 2® Bedside Monitor
- 1.3.8 Spectrum™ Bedside Monitor
- 1.3.9 Spectrum OR™ Bedside Monitor
- 1.3.10 Panorama Telepack
- 1.3.11 Network Printer
- 1.4 Troubleshooting
- 2.0 Getting Started
- 3.0 Alarm Behavior
- 4.0 Arrhythmia and ST Analysis
- 5.0 Patient Setup Functions
- 5.1 Demographics Tab
- 5.2 Transfer Tab
- 5.3 Discharge Tab
- 5.4 Display Tab
- 5.5 Standby Tab
- 5.6 Wave Gain Tab (Patient)
- 5.7 Print Setup Tab (Patient)
- 6.0 Patient Alarm Functions
- 7.0 Views
- 7.1 Bedside Tab
- 7.2 Trends Tab
- 7.3 Events Tab
- 7.4 Disclosure Tab
- 7.4.1 Accessing the Disclosure Tab
- 7.4.2 Disclosure Tab (Waveforms View)
- 7.4.2.1 Accessing the Waveforms View
- 7.4.2.2 Components of the Waveforms View
- 7.4.2.3 Navigating in the Waveforms View of the Disclosure Tab
- 7.4.2.4 ST Button
- 7.4.2.5 Comments Dialog Box (Optional)
- 7.4.2.6 Sidebar Buttons
- 7.4.2.7 Printing a Full Disclosure Zoom In Report from the Waveforms View
- 7.4.2.8 Troubleshooting
- 7.5 ST Tab
- 8.0 Report Functions
- 8.1 Patient Reports Tab
- 8.2 System Reports Tab
- 8.3 Print Status Tab
- 8.4 Additional Reports
- 9.0 System Setup Functions
- 9.1 Parameter Color Tab
- 9.2 Touch Screen Tab
- 9.3 Installation Setup Tab
- 9.4 Care Group Tab
- 9.5 Volume Tab
- 9.6 System Alarms Tab
- 9.7 Passwords Tab
- 9.8 Equipment Setup Tab
- 9.9 More Tab
- 9.10 Previous Tab
- 9.11 Date/Time Tab
- 9.12 Print Setup Tab (System)
- 9.13 Unit Priorities Tab
- 9.14 Unit Choices Tab
- 9.15 Wave Gain Tab (System)
- 9.16 Wireless Tab
- 10.0 Telepack (Optional)
- 11.0 User Maintenance
- 12.0 Parameter Specifications
- 12.1 Parameter Specifications
- 12.1.1 Heart Rate: HR
- 12.1.2 Respiration: Resp
- 12.1.3 NIBP
- 12.1.4 SpO2
- 12.1.5 Nonin SpO2
- 12.1.6 IBP: mmHg
- 12.1.7 CO2
- 12.1.8 O2
- 12.1.9 INSP/ET O2
- 12.1.10 N2O
- 12.1.11 Anesthetic Agents: Iso, Enf, Des, Sev, Hal, Agent
- 12.1.12 Premature Ventricular Contraction (PVC)
- 12.1.13 Temp: T1,T2, Delta T, and T Blood
- 12.1.14 ST
- 12.1.15 Cardiac Output
- 12.1.16 Cardiac Index
- 12.1.17 Pulmonary Artery Wedge Pressure (PAWP)
- 12.1.18 Battery Percentage
- 12.2 Invalid Data Display
- 12.1 Parameter Specifications
- 13.0 Alarms and Events
- 14.0 Network Tab (Optional)
- 14.1 Network Tab
- 14.1.1 Accessing the Network Tab
- 14.1.2 Sorting the Contents of the Network Tab List Box
- 14.1.3 Assigning Patient Tiles in the Network Tab
- 14.1.4 Removing a Patient Tile Assignment in the Network Tab
- 14.1.5 Removing All Patient Tile Assignments in the Network Tab
- 14.1.6 Refreshing the Network Tab List Box
- 14.1.7 Troubleshooting
- 14.1 Network Tab
- 15.0 Paging Tab (Optional)
- 15.1 Paging Password Tab
- 15.2 Paging Tab
- 15.2.1 Accessing the Paging Tab
- 15.2.2 Adding a Pager
- 15.2.3 Assigning a Patient Tile
- 15.2.4 Adding All Patient Tiles to a Pager
- 15.2.5 Removing a Patient Tile Assignment
- 15.2.6 Removing All Patient Tiles from a Pager
- 15.2.7 Editing a Pager
- 15.2.8 Removing a Pager
- 15.2.9 Resending a Failed Page
- 15.2.10 Generating a Paging Report
- 15.2.11 Troubleshooting
- 17.0 Glossary
Panorama™ Operating Instructions 0070-10-0690-01 11 - 5
User Maintenance Telepack User Maintenance
• Soap and water
• Glutaraldehyde 2% (for example, Cidex)
• Dilute chlorine bleach (sodium hypochlorite), 10% solution, freshly made in past 24 hours
• Ethyl Alcohol 30% (for example, green soap tincture)
To clean the ECG lead sets, remove them from the Telepack, and follow these guidelines:
• Use the cleaning solution to wipe and rinse the lead set.
• The cleaning solution should be wiped away with a cloth moistened with distilled water.
• A lint free cloth should be used to dry the lead set thoroughly.
WARNING: The Telepack should not be used adjacent to or stacked with
other equipment. If adjacent or stacked use is necessary, the
Telepack should be observed to verify normal operation in
the configuration in which it will be used.
Sterilization of the Lead Sets
CAUTION: Prior to sterilization, lead sets must be removed from the
Telepack. The Telepack must not be sterilized.
Use EtO to sterilize the leadsets. The sterilization should be performed by following the
hospital-approved procedure for EtO sterilization, such as those recommended by AAMI. The
leadsets remain effective up to 10 exposures to EtO sterilization cycles.
NOTE: Damage to the lead sets will result if steam or gamma
radiation is used for the sterilization process.
11.2.3 Cleaning the SpO
2
Module
In the event that the SpO
2
Module becomes soiled, a cleaning plug (P/N 0003-00-0055)
should be used to cover the serial connector on the SpO
2
Module before it is cleaned.
CAUTION: Datascope requires that the designated cleaning plug (P/N
0003-00-0055) be used when cleaning and disinfecting the
Nonin SpO
2
module. If fluids enter the device, it may affect
the functionality of the module or render it unusable.
Approved Cleaning Solutions
Please refer to the approved cleaning solutions described in ‘‘Cleaning and Disinfecting the
Telepack Device’’ on page 11-3.
Inserting the Cleaning Plug
1. Carefully cover the connector by aligning the cleaning plug with the lower portion of the
SpO
2
Module.