Panorama™ Operating Instructions 0070-10-0690-01 10 - 17
Telepack (Optional) Troubleshooting
10.7 Troubleshooting
This section lists some of the potential messages and issues that may occur while using a
Telepack device.
MESSAGE/ISSUE * REASON SOLUTION
Telepack/LED’s not turning on Batteries are not installed
correctly, batteries are
depleted, or batteries
corroded.
Visually inspect the battery
compartment for any foreign
object.
Replace the batteries.
Telepack not transmitting data
to the Panorama Central
Station.
Batteries are discharged Press the CHECK button to
verify that the batteries are
depleted. Replace as
necessary.
Flashing Battery Status LED
indicator light
Batteries in the Telepack are
weakening
Replace batteries
The message Error: Could
not receive data is
displayed, while attempting to
program the Telepack
Telepack is not connected
properly, or the Panorama
Central Station is unable to
detect the Telepack.
Verify that the service mode
cable is properly connected to
the device.
Retry programming.
Check batteries.
Unlit ECG Lead LED indicator
light
Telepack has no lead
connectivity.
Connect the electrode.
Check batteries.
Communications Lost Telepack’s radio signal is out of
range with the wireless
network
Verify the Telepack is within
100 feet of a network access
point.
Check batteries.
Flashing Link Status LED
indicator light
Telepack is not communicating
with the Panorama Central
Station
Verify that Device ID is
registered on the Panorama
Central Station.
Verify that Panorama Central
Station is functioning.
SpO
2
waveform not displayed
at the Panorama Central
Station patient tile
SpO
2
waveform has not been
selected in the Display tab.
Go to the Display tab and
select the SpO
2
waveform from
the Parameter list to display in
the Waveform 2 position. For
additional information
regarding the Display tab,
refer to ‘‘Configuring the
Display Tab’’ on page 5-45.
SpO
2
module is not properly
connected to the Telepack.
Power down the Telepack and
reconnect the SpO
2
module to
the Telepack.
Battery cap on the Telepack
was not secured.
Verify that the Telepack battery
cap is secured.
SpO
2
Check Sensor An SpO
2
sensor issue was
detected.
Select the digital tile or
waveform tile area, or select
the View button to
acknowledge the technical
event.
* Messages are shown in all bold text.