User Manual

Table Of Contents
Panorama™ Operating Instructions 0070-10-0690-01 10 - 9
Telepack (Optional) Operating Instructions
FIGURE 10-5 Wireless Tab (Telepack-2.4)
6. Select the Read sidebar button. The information from the Telepack device is displayed
in the appropriate fields.
7. For Telepack-2.4 devices, enter the appropriate information in the following fields:
Telepack IP, Subnet Mask, Network Name, and Encryption Key.
8. For Telepack-2.4, devices, select the Security Selection and Lead Selection.
9. Select the Program sidebar button. A dialog box appears.
10. Select the Yes button to confirm. The message Wireless device programmed
successfully appears is displayed confirming that the Telepack was programmed
correctly.
NOTE: If the Telepack device is not connected properly or if the
Panorama Central Station is unable to detect the device, the
message, “Error: Could not receive data” is displayed at the
Panorama Central Station.
10.4.1.3 Entering the Telepack to the Equipment List
The Telepack device should be entered to the Panorama Central Station Equipment List using
the Equipment Setup tab. For additional information about entering a Telepack device to
the Panorama Central Station, refer to “Entering New Equipment” on page 9-34.
The Device ID is found on the back of the Telepack device. After assigning the device to the
Panorama Central Station, the Telepack is ready for patient use.
10.4.1.4 Telepack Power Up Sequence
The Telepack goes through a power up sequence each time the batteries are inserted in the
device. The power up sequence includes a hardware and software check, an ECG Lead
connection status test, a link status test, and a battery status test. A successful power up
sequence is indicated when the LED indicator lights illuminate for approximately 10 seconds.
NOTE: If there are less than two (2) hours of battery life remaining,
the Telepack will not start.
For additional information on the status tests performed by the Telepack during the start up
sequence, refer to the “LEAD SELECT Button” on page 10-3.
After the Telepack device powers up, it is ready to be programmed and assigned to a
patient.