User Manual

Table Of Contents
Wireless Tab System Setup Functions
9 - 66 0070-10-0690-01 Panorama™ Operating Instructions
9.16.3 Entering Wireless Settings (Telepack-608)
This section outlines the settings available in the Wireless tab. The settings in the Wireless
tab are shown by selecting the Read button and configured by selecting the Program
button.
NOTE: Programming and assigning of the Telepack must be
performed by authorized service personnel only.
The fields shown in these images are described on the following pages..
FIGURE 9-47 Wireless Tab after Telepack-608 connected
FIGURE 9-48 Wireless Tab after 608 Instrument Radio connected
Device ID
The Device ID field displays the identification number that was assigned to the connected
wireless device. A Device ID has a maximum of 8 alphanumeric characters. Data is shown in
the Device ID field when the Wireless tab’s Read sidebar button is selected. If the read was
not successful the Device ID field will remain blank. To view the connection status of the
wireless device check the message in the Connect field. The Device ID field cannot be edited
and has a factory default setting of blank.
The Device ID for a monitoring device is set up in the Equipment List tab. For additional
information on setting up a Device ID, refer to the ‘‘Equipment Setup Tab’’ on page 9-33.