User Manual

Table Of Contents
Panorama™ Operating Instructions 0070-10-0690-01 1 - 13
System Overview Supported Devices
1.3.10 Panorama Telepack
The Panorama Telepack is a portable, wireless patient telemetry unit that can operate for an
extended period of time on a disposable battery source. It can measure 3 or 5 Lead ECG
and can also report on SpO2 data when the optional SpO2 module is attached. Telepack
models are available for transmission at either 2.4 GHz (802.11 b/g) or 608MHz (WMTS
band), depending on the region of use. Unless otherwise specified, “Telepack” refers to the
Panorama Telepack-2.4 (2.4 GHz) and Telepack-608 (608 MHz) models.
The Telepack provides basic remote ECG monitoring of a patient, which includes heart rate,
ST, PVC, and arrhythmia. The Telepack sends data to the Panorama Central Station for
analysis and storage. The type of data that can be collected by a Telepack is dependent on
the type of lead wire set and options configured for the connected patient.
Panorama Telepack-2.4
The Telepack-2.4 utilizes the 2.4 GHz wireless band (802.11 b/g). The Telepack-2.4 does
not require a Panorama Server. It communicates directly to the appropriate Panorama
Central Station via a wireless Access Point network installed throughout the specified
coverage area. These Access Points interface with the Panorama Central Station using
Ethernet components.
Panorama Telepack-608
The WMTS Telepack utilizes the 608MHz wireless bands using a frequency division multiple
access method. The Telepack has the capability to transmit to the Panorama Server through
an antenna network installed throughout the specified coverage area.
Panorama Telepack Nonin SpO
2
Module
The Telepack may also monitor SpO
2
if the Nonin SpO
2
module was purchased and is
connected to the Telepack device. The module measures functional oxygen saturation of
arterial hemoglobin (% SpO
2
) and pulse rate (bpm) for adult patients.
NOTE: The Panorama Central Station is used to program,
configure, monitor and collect data for the Panorama
Telepacks.
Parameters/Waveforms
PARAMETERS SETTINGS
PANORAMA
ALARMS
ECG/Heart Rate Waveform leads I, II, III are available with a 3-wire
lead set.
Low Heart Rate
High Heart Rate
Waveform leads I, II, III, aVR, aVL, aVF, V are
available with a 5-wire lead set.
ST ST analysis is enabled ST Single
ST Dual
*) Alarm only available when the Telepack with Nonin SpO
2
module is connected