Panorama™ Operating Instructions 0070-10-0690-01 9 - 51
System Setup Functions Unit Choices Tab
FIGURE 9-40 Unit Choices Tab (ViewStation and WorkStation)
9.14.2.1 Setting the Unit Choices (Central Station)
This section outlines selecting system defaults for pacemaker settings, order of demographic
data and units of measurement.
1. Select the Discharge at Bedside button to set the system default for allowing patient
discharge from the central station.
Discharge at Bedside options include ON and OFF. The factory default setting is OFF.
•The ON option indicates that Discharge at Bedside is enabled.
•The OFF option indicates that Discharge at Bedside is disabled.
NOTE: The Discharge at Bedside button only affects bedside
monitors. This button setting has no effect when a Telepack
device is in use.
NOTE: Changing the Discharge at Bedside option will take affect
the next time the Discharge tab is entered.
2. Select the Pacer Enhancement button to set the system default setting for Pacer
Enhancement. Pacer enhancement will visually mark pacemaker-induced ECG
complexes with a colored vertical spike.
The Pacer Enhancement settings for a specific patient can be changed in the Pacer
Settings dialog box in the Demographics tab. Pacer Enhancement options include
ON and OFF. The factory default setting is ON.
•The ON option indicates that Pacer Enhancement is enabled.
•The OFF option indicates that Pacer Enhancement is disabled.
NOTE: Changing the Pacer Enhancement option will affect all
newly admitted patients and will have no affect on patients
currently being monitored.
NOTE: The Pacer Enhancement option is not available at the
ViewStation or the WorkStation.