9 - 28 0070-10-0690-01 Panorama™ Operating Instructions
System Alarm Limits
Select the System Alarm Limits sidebar button to display the System Alarm Limits
tab. For additional information regarding System Alarm Limits, refer to the ‘‘System Alarms
Tab’’ on page 9-14.
Normal Screen
Select the Normal Screen sidebar button to return the display to the full screen mode, and
close all tabs, message boxes, and dialog boxes.
9.6.4.4 Troubleshooting
This section lists some of the potential messages and issues that may occur while using the
System Alarm Responses tab.
MESSAGE/ISSUE * REASON SOLUTION
The System Setup menu
button is disabled.
The System Setup function is
already open on another
Panorama display and cannot
be opened on multiple
displays.
Close the System Setup
function on the Panorama
display where it is open.
The Page On Alarm button
is not displayed.
The central station where the
patient tile resides does not
have the Paging option
enabled.
If necessary, enable the Paging
option.
The System Alarm Limits
and System Alarm
Responses tab is not
available at the ViewStation or
WorkStation.
System alarm limits and
responses must be changed at
the central station.
Change the alarm limits and
responses from the central
station.
* Messages are shown in all bold text.