User Manual

Table Of Contents
System Alarms Tab System Setup Functions
9 - 16 0070-10-0690-01 Panorama™ Operating Instructions
3. Select the Technical Events Sounds toggle button to enable or disable sound during
a technical event. Technical Events Sounds options include ON and OFF. The
factory default setting is ON.
•The ON option indicates that Technical Events Sounds will be heard during a
technical event.
•The OFF options indicates that Technical Events Sounds will not be heard during
a technical event.
NOTE: Changes to the Technical Events Sounds option will affect all
patients within ten (10) seconds of the change.
4. Select the Apnea Latching toggle button to either enable or disable the apnea
latching function. Apnea latching is the ability to set Apnea alarms to continue the alarm
behaviors until the alarm is acknowledged by the user. Apnea latching options include
ON and OFF. The factory default setting is ON.
•The ON option indicates that Apnea Latching will be enabled.
•The OFF options indicates that Apnea Latching will not be enabled.
NOTE: Changes to the Apnea option will affect all new Apnea
alarms.
5. Select the Paging Delay button to select a paging delay for the system. This button will
only be available when the paging option has been enabled. The Paging Delay options
include 5 seconds, 10 seconds, 15 seconds, 20 seconds, 25 seconds, and 30 seconds.
The factory default setting is 10 seconds.
NOTE: Changing the Paging Delay will only be applied to newly
admitted patients and will not affect patients currently
being monitored or to currently alarming parameters.
NOTE: This option will only be displayed if the Paging option is
enabled at the central station.
6. Select the Password Protection toggle button to either enable or disable the
password protection for the Patient Alarms tab in the Panorama Central Station or
the WorkStation. Password Protection options include ON and OFF. The factory
default setting is ON.
•The ON option indicates that Password Protection will be enabled when trying to
access the Patient Alarms tab.
•The OFF options indicates that Password Protection will not be enabled when
trying to access the Patient Alarms tab.
NOTE: Changes to the Password Protection option will affect all
patients upon entering the Patient Alarms tab.
7. Select the Suspend Alarms toggle button to either enable or disable the ability to
suspend all, all non-lethal, or disable the ability to suspend any alarms. Suspend Alarms
options include All, OFF, and Non-Lethal. The factory default setting is All.
•The All option indicates that the Suspend All Alarms and the Suspend
Non-Lethal sidebar button will be available in the Patient Alarms tabs.
•The Non-Lethal option indicates that only the Suspend Non-Lethal sidebar
button will be available in the Patient Alarms tabs.
•The OFF option indicates that neither the Suspend All Alarms nor the Suspend
Non-Lethal sidebar buttons will be accessible in the Patient Alarms tab.