9 - 6 0070-10-0690-01 Panorama™ Operating Instructions
FIGURE 9-4 Recalibration Screen (Part II)
9.2.2 Sidebar Button - Normal Screen
Select the Normal Screen sidebar button to return the display to the full screen mode, and
close all tabs, message boxes, and dialog boxes.
9.2.3 Troubleshooting
This section lists some of the potential messages and issues that may occur Touch Screen
tab.
Contact your local Datascope Service Representative if recalibration of the touch screen does
not improve the performance of the touch screen display.
MESSAGE/ISSUE * REASON SOLUTION
Recalibration settings not
accepted
The Recalibration screen timed
out because the first target was
not touched within 30 seconds.
Try to recalibrate again and
touch the first target before the
30 second time out.
Recalibration steps are
repeated
The button was
selected in error.
Recalibrate again and if the
targets moved as they were
selected, select the
button.
Touch Screen OFF The Touch Screen On button
was selected.
Use the mouse to navigate. If
touch screen functionality is
needed, press the Touch
Screen OFF button again to
reactivate the touch screen.
* Messages are shown in all bold text.