User Manual

Table Of Contents
Additional Reports Report Functions
8 - 50 0070-10-0690-01 Panorama™ Operating Instructions
This report shows a cursor line in the graph for the Cursor Time posted on the report.
The data plotted in the Graphic Trend Report reflects the trend data records that were
collected based on the Trend Display interval configuration.
The report shows one plotted point for each parameter sensor.
Dual sensor parameters (such as INSP/ET CO2) are plotted individually then connected
with a line (for example, inspired and end tidal values).
Multiple sensor parameters (such as NIBP) are plotted individually then the systolic and
diastolic points are connected and a single point is shown for the Mean measurement.
8.4.5.4 Footer
The Graphic Trend Report contains a footer at the bottom of every page. The footer includes:
The date and time of the print request. The format displayed is taken from the System
Date/Time tab.
A centered page number.
The name of the Panorama Central Station from which the report was generated.
8.4.5.5 Printing a Graphic Trend Report
The Graphic Trend Report is printed in landscape orientation, using any of the following:
Select a valid plot point, then select the Print menu bar button from the Trends
(Graphic) tab.
From the bedside monitor, verify that Select Printer option is configured to print
remotely, and that the trends graphic option is active, then press the PRINT TREND
button.
Requesting this report posts a print job in the system Print Status tab. For additional
information, refer to the “Print Status Tab” on page 8-38.