User Manual

Table Of Contents
Panorama™ Operating Instructions 0070-10-0690-01 7 - 61
Views ST Tab
The Config ST tab at the central
station was closed and all
changes were discarded.
The ST Menu was opened at
the bedside monitor.
Use the bedside monitor to
make the ST configuration
changes.
Close the ST Menu at the
bedside monitor and then
reenter the Config ST tab at the
central station.
The Config ST sidebar button
is disabled.
The These settings are
currently being edited
remotely message is
displayed.
Select the OK button to close
the message box and enable
the sidebar button.
Close the ST Menu at the
bedside device.
MESSAGE/ISSUE * REASON SOLUTION
* Messages are shown in all bold text.