Panorama™ Operating Instructions 0070-10-0690-01 7 - 59
Views ST Tab
Normal Screen
Select the Normal Screen sidebar button to return the display to the full screen mode, and
close all tabs, message boxes, and dialog boxes.
7.5.2 Generating an ST Report
To review the current and reference digital values with the associated ST templates for the
selected patient tile, generate an ST Report. This report can be generated directly from the ST
tab by selecting the Print menu bar button.
NOTE: The Print button is not available when using a WMTS 608
wireless bedside device.
For additional information regarding the ST Report, refer to the ‘‘ST Report’’ on page 8-14.
7.5.3 Troubleshooting
This section lists some of the potential messages and issues that may occur while using the ST
tab.
MESSAGE/ISSUE * REASON SOLUTION
No Patient Selected! is
displayed in the ST tab
demographics text box
This message appears in the
demographic field when an
empty patient tile has been
selected.
Select an active patient tile.
No License is displayed in ST
tab status text box
The monitoring device does not
have an ST license or ST has
been disabled.
Obtain an ST license and/or
enable ST at the monitoring
device.
Disabled is displayed in ST
tab status text box
ST is disabled at the monitoring
device.
Enable ST.
Relearning is displayed in ST
tab status text box
ST was just enabled at the
monitoring device.
Wait for a successful learn
phase to complete.
Real Time is displayed in ST
tab status text box
ST is enabled and a successful
learn phase has been
completed.
No solution required.
Patient in Standby is
displayed in ST tab status text
box
The selected patient tile is in a
standby state.
The message will be removed
when monitoring is resumed.
Communications Lost is
displayed in the ST tab status
text box
The central station is in a
Communications Lost state.
The message will be removed
when communications is
restored.
“---” displays in ST template ST data displays as invalid
because monitor may be in a
learning phase or patient is in
an ECG rhythm that does not
allow for ST monitoring (V-Fib,
Asystole, etc.)
Wait until learn phase is
complete and/or re-establish
normal ECG.
ST templates are present but
contain no ST data.
Monitoring device is in relearn
mode.
Wait for Real Time to display
in ST tab status text box.
* Messages are shown in all bold text.