Panorama™ Operating Instructions 0070-10-0690-01 7 - 39
Views Events Tab
ST button is disabled ST was never enabled for the
selected patient tile.
If the patient tile requires ST analysis,
and there is an ST license available,
enable ST from the bedside monitor.
Lead sidebar button is
disabled
The selected device is a:
Telepack device using a 3
or 5-wire lead
Telepack with Nonin
SpO
2
module attached
with a 3-wire lead set
No solution required.
The All ECG sidebar button is
latched.
Select the All ECG sidebar button to
enable the Lead sidebar button.
CO/CI data is not
displayed in Waveforms
view but is displayed in
Events list.
Currently, CO/CI waveform
data is not displayed in the
Waveforms view.
No solution required.
No more events found The colored vertical cursor is
positioned at the first or the
last event in the Events
database.
No solution required.
Comments sidebar
button is not displayed in
sidebar.
The Paging option is not
present on the central station
licensing key.
Obtain a licensing key that contains
the Paging option.
Comments sidebar
button is disabled.
No paging information was
sent for the selected event.
If a page should have been sent for
the selected event type, verify the
Page On Alarm settings in the Alarm
Responses tab.
MESSAGE/ISSUE * REASON SOLUTION
* Messages are shown in all bold text.