User Manual

Table Of Contents
Panorama™ Operating Instructions 0070-10-0690-01 7 - 37
Views Events Tab
Done Button
Select the Done button to close the Waveforms view ST dialog.
7.3.5.4 Comments Dialog Box (Optional)
With the addition of system-wide paging, a paging notification can be sent to clinicians to
alert them that a specific event type has occurred. If a page was sent for a selected event in
the Waveforms view, an asterisk (*) will appear beside the event in the event description
area (shown in FIGURE 7-18). The paging information related to the event can be viewed by
selecting the Comments sidebar button. The Comments dialog box will display either a
Page Sent or a Page Failed message. If the selected event does not have an asterisk
beside the event description, this indicates that a page has not been sent for the selected
event. In this case, the Comments sidebar button will be disabled.
7.3.5.5 Sidebar Buttons
The Waveforms view may contain up to six sidebar buttons.
View List
Select the View List sidebar button to return to the Events list.
Lead
Select the Lead sidebar button to cycle through each of the waveforms in the ECG lead
ordered set. The Lead sidebar button is not available when using a:
Telepack device with a 3 or 5-wire lead set
Telepack with Nonin SpO
2
module with a 3-wire lead set
All ECG
Select the All ECG sidebar button to access the All ECG mode of the Events tab. When the
All ECG button is white, this indicates the Events tab is in the All ECG mode. The
Waveforms view remains in the All ECG mode until the All ECG sidebar button is
selected again. The All ECG sidebar button is not available when using a:
Telepack device with a 3 or 5-wire lead set
Telepack with Nonin SpO
2
module with a 3-wire lead set
NOTE: If a 5-wire or 12-lead set was previously used for the
selected patient, the All ECG sidebar push button will
always be enabled.
Measure
Select the Measure sidebar button to activate the Measure dialog box (shown in
FIGURE 7-3). The Measure dialog contains reference and measurement lines to calculate the
time and amplitude of an ECG waveform. For instructions on the use of the Measure button,
refer to ‘‘Performing ECG Waveform Measurements’’ on page 7-5.
NOTE: The Print menu bar will be enabled when the Measure
dialog box is open.