User Manual

Table Of Contents
Panorama™ Operating Instructions 0070-10-0690-01 7 - 17
Views Trends Tab
Graphic
Select the Graphic sidebar button to open the Graphic view of the Trends tab. For
additional information regarding the Graphic view, refer to "Accessing the Graphic View"
on page 7-19.
Trend Display
This sidebar button was previously described on page 7-15.
Normal Screen
Select the Normal Screen sidebar button to return the display to the full screen mode, and
close all tabs, message boxes, and dialog boxes.
7.2.2.6 Troubleshooting
This section lists some of the potential messages and issues that may occur while using the
List view of the Trends tab.
MESSAGE/ISSUE * REASON SOLUTION
There is no data in the List
view of the Trends tab.
A patient has not been
admitted to the selected tile.
Admit a patient to the selected tile or
choose a tile that already has an
admitted patient.
The Trends tab was selected
before the first interval was
reached or before the first
NIBP measurement occurred.
Wait until the first interval is reached
or until the first NIBP measurement
occurs.
A column of parameter
data is not displayed.
If the column of parameter
data exists, it has not been
scrolled into view on the
display.
Use the scroll bar located along the
bottom of the Trend list to scroll to the
desired parameter column.
The parameter has not been
monitored for this patient.
Connect the sensor(s) for the
parameter to the patient.
No more data found
(Page Down Button)
There is no data in the Trend
database because the
Trends tab was selected
before the first interval was
reached or before the first
NIBP measurement occurred.
Wait until the first interval is reached
or until the first NIBP measurement
occurs.
The last row in the list is
already highlighted.
Select the OK button to close the
message box and return to the List
view of the Trends tab.
No more data found
(Page Up Button)
There is no data in the Trend
database because the
Trends tab was selected
before the first interval was
reached or before the first
NIBP measurement occurred.
Wait until the first interval is reached
or until the first NIBP measurement
occurs.
The first row in the list is
already highlighted.
Select the OK button to close the
message box and return to the List
view of the Trends tab.
* Messages are shown in all bold text.