7 - 6 0070-10-0690-01 Panorama™ Operating Instructions
2. Adjust the location of the vertical reference line as follows:
For coarse adjustments, select the desired location on the touch screen. The reference
line will move to that point.
For fine adjustments, use the left/right arrow buttons to position the reference line at
the desired point.
3. Once the vertical reference line appears in the desired location, select the Calculate
button. The measurement populates the text box beside the selected interval.
If another vertical measurement is desired, select another interval button.
To delete a measurement, select the interval button that needs to be to reset, then select
the Clear button.
Horizontal Measurements (Amplitude)
1. Select the ST interval. A dashed horizontal reference line is placed above the solid
horizontal reference line.
NOTE: Adjustments can be made to both the horizontal solid
reference line and/or the horizontal dashed reference line.
Selecting the interval button a second time toggles between
the two reference lines. When a reference line is active, it
blinks/flashes on the screen.
2. Adjust the location of the dashed, horizontal reference line as follows:
For coarse adjustments, select the desired location on the touch screen. The reference
line will move to that point.
For fine adjustments, use the up/down arrow buttons to position the reference line at
the desired point.
3. Once the dashed, horizontal reference line appears in the desired location, select the
Calculate button. The measurement populates the text box beside the ST interval.
If another horizontal measurement is desired, select another interval button.
To delete the ST measurement, select the ST interval button, then select the Clear
button.
7.1.2.4 Sidebar Buttons
The following sidebar buttons are used to expand the functionality of this view.
Lead
Select the Lead sidebar button to cycle through each of the waveforms in the ECG ordered
set. The Lead sidebar button is not available when using a:
Telepack device with a 3 or 5-wire lead set
Telepack with Nonin SpO
2
module with a 3-wire lead set