Panorama™ Operating Instructions 0070-10-0690-01 7 - 5
Views Bedside Tab
7.1.2.2 Navigating in the All Waveforms View
If the waveforms exceed the capacity of the viewable screen, use the navigational buttons to
locate the desired waveform. The buttons are described as follows:
The up arrow button moves the group of displayed waveforms up one position within
the ordered set of waveforms
The down arrow button moves the group of displayed waveforms down one position
within the ordered set of waveforms
7.1.2.3 Performing ECG Waveform Measurements
ECG waveform measurements for time and amplitude can be performed in the All
Waveforms view by using the Measure dialog box (shown in FIGURE 7-3).
FIGURE 7-3 Measure Dialog Box
The Measure dialog box can only be accessed once the Freeze sidebar button has been
selected. When the Freeze sidebar button is selected, the All ECG sidebar button is
replaced with the Measure sidebar button.
When the Measure dialog box is first accessed, it contains five interval buttons. Once an
interval button is selected, two changes occur:
Navigational buttons and the Clear and Calculate buttons are displayed in the
Measure dialog box
Reference and measurement lines are displayed in the Bedside tab waveform area
The navigational buttons displayed depend on the selected interval. The left and right arrow
buttons are available when using the PR, QRS, QT, and R to R intervals. The up and down
arrow buttons are available when using the ST interval.
Vertical Measurements (Time)
1. Select either the PR, QRS, QT, or R to R intervals. A dashed vertical reference line is
placed to the left of the solid vertical reference line.
NOTE: Adjustments can be made to both the solid vertical reference
line and/or the vertical dashed reference line. Selecting the
interval button a second time toggles between the two
reference lines. When a reference line is active, it
blinks/flashes on the screen.