User Manual

Table Of Contents
Alarm Limits Tab (Patient) Patient Alarm Functions
6 - 6 0070-10-0690-01 Panorama™ Operating Instructions
2. Use the scroll bar or the arrow buttons to locate the parameter to be modified.
Double arrow buttons display the next/previous page of parameter slider controls
(until the first/last page is reached).
Single arrow buttons display the next/previous slider control (until the last control is
shown).
3. Set the High Alarm Limit button for the selected parameter. This button is either
labeled OFF or contains an alarm limit value. A high alarm limit is set using the up/
down arrow buttons on the parameter’s vertical scroll bar or by moving the alarm’s
slider control.
4. Set the Low Alarm Limit button for the selected parameter. This button is either
labeled OFF or contains an alarm limit value. A low alarm limit is set using the up/down
arrow buttons on the parameter’s vertical scroll bar or by moving the alarm’s slider
control.
Alarm limit adjustments made at the Panorama Central Station are automatically sent to the
monitoring device.
For additional information regarding the default alarm limit ranges, refer to ‘‘Alarms and
Events’’ on page 13-1.
Auto Set Alarms Button
The Auto Set button automatically recalculates alarm limit settings to the percent indicated
(shown in Table 6-1) above and below the current measured value.
NOTE: The Auto Set button does not affect the SpO
2
, ST, PVC,
Arrhythmia Run or Apnea Delay settings.
NOTE: The Auto Set button is not available at the ViewStation.
TABLE 6-1 Auto Set Percent Indicator
ALARM
PERCENT INDICATOR
(APPROXIMATE)
Heart Rate 30%
Respiration 30%
IBP Systolic 20%
IBP Diastolic 20%
IBP Mean 20%
NIBP Systolic 20%
NIBP Diastolic 20%
NIBP Mean 20%
ET CO
2
(mmHg) 20%
INSP CO
2
(mmHg) 20%
ET CO
2
(%) 20%
INSP CO
2
(%) 20%
ET CO
2
(kPa) 20%
INSP CO
2
(kPa) 20%
Temp (F°) 3%
Temp (C°) 3%