Panorama™ Operating Instructions 0070-10-0690-01 5 - 53
Patient Setup Functions Wave Gain Tab (Patient)
FIGURE 5-34 Wave Gain Tab (ViewStation and WorkStation)
5.6.2 Changing a Patient’s Wave Gain Settings
This section outlines entering and modifying a patient’s wave gain settings using the Wave
Gain tab.
ViewStation and WorkStation patient tiles will initially display the same wave gain settings of
the associated patient tile at the host central station. For additional information regarding the
transfer of wave gain settings from the host central station, refer to ‘‘Wave Gain Tab
(Patient)’’ on page 5-52.
1. Choose a patient tile by selecting its digital data or waveform area.
2. In the Wave Gain tab, select the ECG mm/mV button until the desired wave gain
displays. The settings available for the ECG wave gain include 2.5, 5.0, 10.0, 20.0,
and 30.0 mm/mV.
3. Select the IBPx mmHg buttons until the desired wave gain displays. The settings
available for the IBPx wave gain include 0-20, 0-40, 0-80, 60-140, 0-160, 0-225,
0-320, and (-10) -10 mmHg.
NOTE: If the IBPx button(s) have been labeled at the bedside
monitor, the same label(s) will be displayed on the
individual wave gain button(s).
4. Select the Resp Gain button until the desired wave gain is displayed. The settings
available for the resp gain wave gain include 1, 2, 3, 4, and 5.
5. Select the CO2 (%, kPa, mmHG) button until the desired wave gain is displayed.
The settings available for the CO
2
% wave gain include 5.0, 7.5, and 10.0 %.
The settings available for the CO
2
kPa wave gain include 5.0, 7.5, and 10.0 kPa.
The settings available for the CO
2
mmHg wave gain include 40, 60, and 100 mmHg.
6. Select the O2 % button until the desired wave gain is displayed. The settings available
for the O
2
wave gain include 18 - 30, 18 - 60, and 18 - 100 %.
7. Select the AGENT % button until the desired wave gain is displayed. The settings
available for the agent wave gains include 0 - 1.0, 0 - 2.5, 0 - 5.0, 0 - 10.0,
0 - 15.0, and 0 - 20.0 %.
8. Select the Done button to accept the wave gain settings shown in the Wave Gain tab.
A patient’s wave gain settings will not be changed unless the Done button is selected.