5 - 48 0070-10-0690-01 Panorama™ Operating Instructions
5.4.4 Sidebar Buttons
The following sidebar buttons are used to expand the functionality of this tab.
Restore Previous Settings
Select the Restore Previous Settings sidebar button to return patient display settings to
those in use when the tab was opened.
Trend Display
This sidebar button was previously described on page 5-46.
Normal Screen
Select the Normal Screen sidebar button to return the display to the full screen mode, and
close all tabs, message boxes and dialog boxes.
5.4.5 Troubleshooting
This section lists some of the potential messages and issues that may occur while using the
Display tab.
MESSAGE/ISSUE * REASON SOLUTION
No Patient Selected! This message is displayed in
the demographic field when
either a patient tile has not
been selected or an empty tile
was selected.
Select an active patient tile.
V2, V3, V4, V5, or V6 was
selected from the Parameter list
but the waveform is not
displayed in patient tile.
A WMTS 608 wireless bedside
device with a View 12 card is
in currently in use. When using
this device type, the V2-V6
waveforms will not be
displayed.
Select a different waveform
from the Parameter list.
Change to a hardwired or 2.4
GHz wireless device.
The parameter tiles in the
Display tab are disabled.
A valid patient tile is not
selected.
Select an active patient tile.
* Messages are shown in all bold text.