Panorama™ Operating Instructions 0070-10-0690-01 5 - 45
Patient Setup Functions Display Tab
5.4.3 Configuring the Display Tab
This section describes how to configure parameter tiles in the Display tab.
ViewStation and WorkStation patient tiles will initially display the same display settings of
the associated patient tile at the host central station. For additional information regarding the
transfer of display settings from the host central station, refer to the ‘‘Trends Tab’’ on page
7-13.
Selecting a Patient’s Parameter Data
1. Choose a patient tile by selecting its digital data or waveform area.
2. Select the parameter tile to be modified. The Parameter list is displayed.
3. From the Parameter list, select the parameter to be viewed or turn the parameter OFF.
NOTE: Digital parameters and the first waveform parameter
cannot be turned off.
NOTE: The OFF option is only available in the Parameter list when
Waveform 2 is selected in the Waveform Display area.
When OFF is selected, the message “Are you sure you want
to turn off the parameter” is displayed.
Digital Parameters include Heart Rate, Respiration, IBP, NIBP, Temp, ST, PVC, SpO
2
,
N
2
O, CO
2
, O
2,
Cardiac Output (CO), Cardiac Index (CI), T Blood, Delta T,
Pulmonary Artery Wedge Pressure (PAWP), Battery Percentage
(for a Telepack device
only), and Agent.
NOTE: The CO/CI, Temp 2, and T Blood parameters are only
available if they are being monitored by the Spectrum/
Spectrum OR External Parameter Module (EPM). These
parameters are not available if they are monitored by an
Edwards Vigilance
®
monitor.
Waveform Parameters include I, II, III, aVR, aVL, aVF, V, V1, V2, V3, V4, V5, and V6.
4. Choose the Select button to highlight the selected parameter. The parameter is shown
in the selected patient tile.
NOTE: A digital parameter can only be shown in one parameter
tile at any given time.
Waveform 2 I, II, III, aVR, aVL, aVF, V, V1, V2, V3, V4, V5, V6, Resp, SpO
2
, IBP1, IBP2,
IBP3, IBP4, CO
2,
O
2
, Agent (Iso, Enf, Sev, Hal, Des)
The ECG waveforms available are dependent on the lead set used. The
waveform chosen for the ECG Lead waveform will not be included in the
parameter list because it cannot be used in both waveform positions.
The remaining parameters available for the second waveform are
dependent on the monitoring device associated with the patient tile.
WAVEFORM
PARAMETER AVAILABLE WAVEFORMS