5 - 44 0070-10-0690-01 Panorama™ Operating Instructions
Waveform Parameter Data
The patient tile is capable of showing up to two waveforms for a patient (shown in
FIGURE 5-28).
NOTE: The initial central station wave default settings come from
the System Wave Default settings, if settings were
configured.
FIGURE 5-28 Display Tab - Waveform Data
FIGURE 5-29 Display Tab - Waveform Data (ViewStation and WorkStation)
The first waveform is always an ECG waveform. The second waveform can be any other
waveform parameter available for the associated device or, if desired, it can be turned off.
Each waveform has its own baseline.
NOTE: Waveforms display in the color that was selected for the
parameter. For additional information, refer to the
‘Parameter Color Tab’’ on page 9-2.
The following table shows the available waveform parameters.
WAVEFORM
PARAMETER AVAILABLE WAVEFORMS
ECG Lead I, II, III, aVR, aVL, aVF, V, V1, V2, V3, V4, V5, V6
The ECG waveforms available are dependent on the lead set used.
ECG waveforms use the Pacer Enhancement and Filter Settings that were
selected in the Demographics tab for the selected patient tile.